ORIGINAL PAPER
Dosages of ionizing radiation during limb diagnostic X-ray examinations
 
More details
Hide details
1
National Institute of Public Health – National Institute of Hygiene / Narodowy Instytut Zdrowia Publicznego – Państwowy Zakład Higieny, Warszawa, Poland (Department of Radiation Protection and Radiobiology / Zakład Higieny Radiacyjnej i Radiobiologii)
 
2
Central Laboratory for Radiological Protection / Centralne Laboratorium Ochrony Radiologicznej, Warszawa, Poland (Department of Radiation Hygiene / Zakład Higieny Radiacyjnej)
 
 
Corresponding author
Marcin Bekas   

National Institute of Public Health – National Institute of Hygiene, Department of Radiation Protection and Radiobiology, Chocimska 24, 00-791 Warszawa, Poland
 
 
Med Pr Work Health Saf. 2016;67(3):321-6
 
KEYWORDS
TOPICS
ABSTRACT
Background: X-ray examination is associated with the patient’s exposure to ionizing radiation. The dose values depend on the type of the medical procedure used, the X-ray unit technical condition and exposure conditions selected by X-ray technicians. The aim of this study has been to assess the entrance surface dose (ESD) values received by patients during the limb X-ray examination. The results should help doctors in making the decision about sending patients for X-ray examination. At the same time the X-ray unit condition and examination method performance are important for the radiological protection of the medical staff. Material and Methods: The study covered the total number of 118 X-ray units located in 56 public healthcare entities and private medical centers in the Masovian Voivodeship. The measurement of the radiation dose rate received by patients was based on our own research procedures. Results: The research has found that there are even more than 10-fold differences in the dose values received by adult patients with several-fold differences in the case of children patients. The broadest dose value range for adult patients was related to femur radiography. The ESD values for this procedure ranged 70.9–765.2 μGy (with the average value of 319.7 μGy). The broadest dose value range for children was related to the knee radiography. The range for children aged 5 years old was 11.8–95.8 μGy (with the average value of 48.9 μGy). Conclusions: It is essential to immediately implement X-ray room working procedures for the purpose of performing diagnostic examinations based on the existing model procedures. Med Pr 2016;67(3):321–326
eISSN:2353-1339
ISSN:0465-5893
Journals System - logo
Scroll to top