ORIGINAL PAPER
Calibration of thermoluminescent detectors in Hp(0.07) units by using an X-ray tube and a 137Cs source
More details
Hide details
1
University of Lodz, Łódź, Poland (Faculty of Physics and Applied Informatics, Department of Nuclear Physics and Radiation Safety)
Online publication date: 2019-09-24
Corresponding author
Małgorzata Wrzesień
University of Lodz, Faculty of Physics and Applied Informatics,
Department of Nuclear Physics and Radiation Safety, Pomorska 149/153, 90-236 Łódź, Poland
Med Pr Work Health Saf. 2019;70(6):669-73
KEYWORDS
TOPICS
ABSTRACT
Background: The method of measuring doses based on the thermoluminescence phenomenon is not an absolute method. For this
reason, to obtain correct results, it is necessary to calibrate detectors in the known radiation field. This paper presents a method for
calibrating thermoluminescent detectors used in the measurement of personal dose equivalents (Hp(0.07)) obtained by nuclear medicine
facility personnel when handling the 99mTc radionuclide. Material and Methods: The authors used self-developed high-sensitivity
thermoluminescent detectors and a HF320C X-ray unit, as well as a rod phantom. Dosimeters were calibrated in accordance
with the ISO 4037-3 standard. During the measurements a vial containing a 99mTc radionuclide with well-known activity was also
used. The energy characteristics were supplemented by using a 137Cs source (irradiator 60Co/137Cs). Results: The value of the calibration
coefficient for 118 keV energy energy was (1.90±0.02)×10−5 mSv/imp. Taking into account the correction factor specified for of
140 keV energy at 0.962, the value of the calibration coefficient for 140 keV energy was determined as (1.83±0.02)×10−5 mSv/imp. Conclusions: Verification of the calibration coefficient determined for 140 keV energy carried out with a vial containing a 99mTc
radionuclide confirmed the correctness of the procedure. Med Pr. 2019;70(6):669–73